Continuous monitoring of electrode-skin impedance mismatch during bioelectric recordings

2008
Article
Résumé

In bioelectric recordings, an electrode-skin impedance mismatch leads to a reduced common-mode rejection ratio (CMRR) of the amplifier. For this reason, the impedance of each individual electrode-skin contact is usually measured prior to a recording. The measurement circuit itself degrades the CMRR and is switched off during the bioelectric recording. In this paper, we present a new method that allows to monitor the electrode-skin impedance in a continuous way without reducing the CMRR of the amplifier. The new method is based on an additional common-mode signal that is superimposed on the bioelectric signal.

À propos de ce résultat
Cette page est générée automatiquement et peut contenir des informations qui ne sont pas correctes, complètes, à jour ou pertinentes par rapport à votre recherche. Il en va de même pour toutes les autres pages de ce site. Veillez à vérifier les informations auprès des sources officielles de l'EPFL.

Graph Chatbot

Chattez avec Graph Search

Posez n’importe quelle question sur les cours, conférences, exercices, recherches, actualités, etc. de l’EPFL ou essayez les exemples de questions ci-dessous.

AVERTISSEMENT : Le chatbot Graph n'est pas programmé pour fournir des réponses explicites ou catégoriques à vos questions. Il transforme plutôt vos questions en demandes API qui sont distribuées aux différents services informatiques officiellement administrés par l'EPFL. Son but est uniquement de collecter et de recommander des références pertinentes à des contenus que vous pouvez explorer pour vous aider à répondre à vos questions.