This paper presents a novel sensor front-end circuit that addresses the issues of 1/f noise and distortion in a unique way by using canceling techniques. The proposed front-end is a fully differential transimpedance amplifier (TIA) targeted for current mode electrochemical biosensing applications. In this paper, we discuss the architecture of this canceling based front-end and the optimization methods followed for achieving low noise, low distortion performance at minimum current consumption are presented. To validate the employed canceling based front-end, it has been realized in a 0.18 mu m CMOS process and the characterization results are presented. The front-end has also been tested as part of a complete wireless sensing system and the cyclic voltammetry (CV) test results from electrochemical sensors are provided. Overall current consumption in the front-end is 50 mu A while operating on a 1.8 V supply.
Paul Joseph Dyson, Lucinda Kate Batchelor, Silvia Schoch
Hubert Girault, Astrid Johana Olaya Avendano, Jorge Gustavo Uranga, Julieta Soledad Riva, Sara Natalia Moya Betancourt
Rosario Scopelliti, Marinella Mazzanti, Ivica Zivkovic, Anne-Sophie Chauvin