Background The identification of specific clot characteristics before mechanical thrombectomy (MTB) might allow the selection of the most effective first-line technique, thus potentially improving the procedural outcome. We aimed to evaluate if the microwire push forces could extrapolate information on clot consistency and extension before MTB, based on clot mechanical properties. Methods We measured in vitro the forces exerted on the proximal extremity of the guidewire during the advancement and retrieval of the guidewire through clot analogs of different compositions. In addition, we analyzed the forces exerted on the guidewire to extrapolate information about the location of the proximal and distal extremities of the clot analogs. Results The maximum forces recorded during the whole penetration phase were significantly different for hard and soft clots (median values, 55.6 mN vs 15.4 mN, respectively; P