Êtes-vous un étudiant de l'EPFL à la recherche d'un projet de semestre?
Travaillez avec nous sur des projets en science des données et en visualisation, et déployez votre projet sous forme d'application sur Graph Search.
The aim of this study is to compare the difference in µ-motion between a custom stem implant and a sps standard stem and assess the importance of patient specificity. A surgeon would implant 3 sps-standar stems and 3 custom stems in 6 cadaveric femurs. The implants will be loaded at different physiological strengths, by means of a mechanical device attached to the distal femoral end and the implant neck. The different deformations induced by the loadings are going to be recorded by computed tomography utilizing a µ-CT SkyScan 1076 in-vivo µCT The student will be responsible for performing the µ-CT scans of the femurs and interpreting the data. Which include: Number of contact points and total contact surface area. Gap sizes (extension and normal distance). Locations of gaps and contact areas. Gap variation as a function of load (µ-motion). Subsidence. Bone quality of each bone sample. Variation of above parameters according to the individuals.
Roland Logé, Christian Leinenbach, Efthymios Polatidis, Markus Strobl, Nikola Kalentics