Are you an EPFL student looking for a semester project?
Work with us on data science and visualisation projects, and deploy your project as an app on top of Graph Search.
The goal of this paper is to prove that a safe and efficient energy transfer is possible between an external transducer located on the patient's skin and a device deeply implanted in the abdomen. An ultrasound propagation model based on the Rayleigh-Sommerfeld diffraction integral is coupled with the data from the Visible Human Project to account for the geometry of the organs in the body. The model is able to predict the amount of acoustic power received by the device for different acoustic paths. The acoustic model is validated by comparison with measurements in water and in heterogeneous liquid phantoms. Care is taken to minimize adverse bioeffects-mainly temperature rise and cavitation in tissues. Simulations based on the bio-heat transfer equation are performed to check that thermal effects are indeed small.
Romain Christophe Rémy Fleury, Janez Rus
Christoph Koch, Peter Lindner, Zilu Tian, Val Tannen