A new laser-Doppler flowmeter at large interoptode spacing, based on single-photon counting (single-photon laser-Doppler flowmetry [SP-LDF]) and allowing assessment of blood flow deep in bone tissue, is proposed and implemented. To exploit the advantages of the new SP-LDF hardware, a dedicated simple and efficient time-domain algorithm has been developed. The new algorithm is based on the zero-order moment of the power density spectrum of the ad hoc prefiltered photoelectric current. The SP-LDF has been validated by Monte Carlo simulations, as well as by experimental measurements on a bone tissue phantom for optical flowmeters and on human. (C) 2014 Optical Society of America
Christian Heinis, Xudong Kong, Jonas Alfred Karl Wilbs, Patrick Andreas Gonschorek, Simon Jan Middendorp, Christina Lamers
Josefine Catharina Pedersen Tratwal