How to prevent venous cannula orifice obstruction during extracorporeal circulation

Julien Favre
2015
Journal paper
Abstract

Venous cannula orifice obstruction is an underestimated problem during augmented cardiopulmonary bypass (CPB), which can potentially be reduced with redesigned, virtually wall-less cannula designs versus traditional percutaneous control venous cannulas. A bench model, allowing for simulation of the vena cava with various affluent orifices, venous collapse and a worst case scenario with regard to cannula position, was developed. Flow (Q) was measured sequentially for right atrial + hepatic + renal + iliac drainage scenarios, using a centrifugal pump and an experimental bench set-up (afterload 60 mmHg). At 1500, 2000 and 2500 RPM and atrial position, the Q values were 3.4, 6.03 and 8.01 versus 0.77*, 0.43* and 0.58* l/min: p