Background: Use of a T2 mapping sequence in addition to the conventional knee MRI protocol increases sensitivity to early cartilage lesions but is time consuming. Purpose: To test the in vitro validity of quantitative data from an accelerated parallel T2 mapping sequence (combined generalized autocalibrating partially parallel acquisition and model-based accelerated relaxometry by iterative nonlinear inversion [GRAPPATINI]) of the knee and to compare in vivo synthetic images generated with this sequence with those generated with conventional morphologic sequences. Materials and Methods: T2 estimations with GRAPPATINI were validated in vitro in comparison with T2 estimations with routine multisection multiecho and reference standard single-section single-echo spin-echo T2 mapping sequences by using a Bland-Altman plot. Synthetic morphologic images (intermediate-weighted sequence, 34-msec echo time; T2-weighted sequence, 80-msececho time) were compared in vivo with corresponding conventional morphologic turbo spin-echo 3-T sequences by three readers in consecutive patients recruited retrospectively from February to May 2018. Synthetic and conventional morphologic images were compared by using rates of interreader agreement, k statistics, and rates of findings. Results: T2 values with GRAPPATINI were accurate compared with those obtained with the reference single-section single-echo sequence,with slight T2 overestimation (2.7 msec). Sixty-one patients (mean age, 43 years +/- 16 [standard deviation]; 32 men) were included. The rate of agreement when one reader used synthetic morphologic images and the other used conventional sequences was not inferior to the rate of agreement when all readers used conventional sequences (upper bounds of 95% confidence intervals of differences between rates of agreement
Anthony Christopher Davison, Timmy Rong Tian Tse
Anthony Christopher Davison, Igor Rodionov