In electrochemistry, the auxiliary electrode, often also called the counter electrode, is an electrode used in a three-electrode electrochemical cell for voltammetric analysis or other reactions in which an electric current is expected to flow. The auxiliary electrode is distinct from the reference electrode, which establishes the electrical potential against which other potentials may be measured, and the working electrode, at which the cell reaction takes place.
In a two-electrode system, either a known current or potential is applied between the working and auxiliary electrodes and the other variable may be measured. The auxiliary electrode functions as a cathode whenever the working electrode is operating as an anode and vice versa. The auxiliary electrode often has a surface area much larger than that of the working electrode to ensure that the half-reaction occurring at the auxiliary electrode can occur fast enough so as not to limit the process at the working electrode.
When a three-electrode cell is used to perform electroanalytical chemistry, the auxiliary electrode, along with the working electrode, provides a circuit over which current is either applied or measured. Here, the potential of the auxiliary electrode is usually not measured and is adjusted so as to balance the reaction occurring at the working electrode. This configuration allows the potential of the working electrode to be measured against a known reference electrode without compromising the stability of that reference electrode by passing current over it.
The auxiliary electrode may be isolated from the working electrode using a glass frit. Such isolation prevents any byproducts generated at the auxiliary electrode from contaminating the main test solution: for example, if a reduction is being performed at the working electrode in aqueous solution, oxygen may be evolved from the auxiliary electrode. Such isolation is crucial during the bulk electrolysis of a species which exhibits reversible redox behavior.
Cette page est générée automatiquement et peut contenir des informations qui ne sont pas correctes, complètes, à jour ou pertinentes par rapport à votre recherche. Il en va de même pour toutes les autres pages de ce site. Veillez à vérifier les informations auprès des sources officielles de l'EPFL.
The course covers the fundaments of bioelectronics and integrated microelectronics for biomedical and implantable systems. Issues and trade-offs at the circuit and systems levels of invasive microelec
This course builds upon the underlying theory in thermodynamics, reaction kinetics, and transport and applies these methods to electrosynthesis, fuel cell, and battery applications. Special focus is p
thumb|upright=1.4|Système à trois électrodes : 1) électrode de travail 2) électrode auxiliaire 3) électrode de référence. Une électrode de travail est une électrode dans un système électrochimique sur laquelle se produit la réaction d'intérêt. L'électrode de travail est souvent utilisée avec une électrode auxiliaire et une électrode de référence dans un système à trois électrodes. Selon que la réaction sur l'électrode est une réduction ou une oxydation, l'électrode de travail est appelée cathodique ou anodique.
A liquid metal electrode is an electrode that uses a liquid metal, such as mercury, Galinstan, and NaK. They can be used in electrocapillarity, voltammetry, and impedance measurements. The dropping mercury electrode (DME) is a working electrode made of mercury and used in polarography. Experiments run with mercury electrodes are referred to as forms of polarography even if the experiments are identical or very similar to a corresponding voltammetry experiment which uses solid working electrodes.
Une électrode de référence est une électrode qui maintient un potentiel pratiquement invariant dans les conditions prévalant dans une mesure électrochimique. Elle permet l'observation, la mesure ou le contrôle du potentiel de l'indicateur (ou de l'essai) ou de l'électrode de travail.On l'utilise en électrochimie pour étudier les propriétés d'oxydoréduction et en chimie analytique pour mesurer les concentrations d'espèces dissoutes dans une solution. Le potentiel de l'électrode de référence ne doit en aucun cas varier pendant la mesure.
Explore la fabrication et le comportement des dispositifs quantiques et nanoinformatiques, en se concentrant sur la conduction moléculaire et les caractéristiques des dispositifs.
Permissible limits of Pb 2+ in drinking water are being reduced from 10 mu gL -1 to 5 mu gL -1 , which calls for rapid, and highly reliable detection techniques. Electrochemical sensors have garnered attention in detection of heavy metal ions in environmen ...
Cell Press2024
, , ,
Anion exchange membrane water electrolysis (AEMWE) offers a green hydrogen production method that eliminates the need for platinum group metals (PGM) as electrocatalysts. This study employs a COMSOL (R) 6.0 model to simulate a 1x1 cm(2) Ni fibre - Raney (R ...
Amsterdam2024
, , , , ,
Dielectric elastomer actuators (DEAs) have raised interest due to their remarkable capabilities in various applications, such as soft robotics, haptic feedback systems, and biomedical devices. To harness the full potential of DEAs, the choice of the electr ...
Society of Photo-Optical Instrumentation Engineers (SPIE)2024