Êtes-vous un étudiant de l'EPFL à la recherche d'un projet de semestre?
Travaillez avec nous sur des projets en science des données et en visualisation, et déployez votre projet sous forme d'application sur Graph Search.
We report on the development of a time-of-flight (ToF) mass spectrometer with a highly efficient electrostatic ion guide for enhancing detectability in ToF mass spectrometry. This 65-cm long ion guide consists of 13 cascaded stages of Einzel lens to collect a large fraction of emitted charges over a wide emission angle and energy spread for time-of-flight measurements. Simulations show that the ion guide can collect 100% of the charges with up to 23° emission half-angle or 30% energy spread irrespective of their specific charge. We demonstrate this ion guide as applied to electrospray ion sources. Experiments performed with tungsten needle electrospraying the ionic liquid EMI- BF4 showed that up to 80% of the emitted charges could be collected at the end of the flight tube. Flight times of monomers and dimers emitted from the needles were measured in both positive and negative emission polarities. The setup was also used to characterize the electrospray from° microfabricated silicon capillary emitters and nearly 30% charges could be collected even from a 40 emission half-angle. This setup can thus increase the fraction of charge collection for ToF measurement and spray characteristics can be obtained from a very large fraction of the emission in real time.
Ivo Fabio Beck, Benjamin Jérémy Laurent Heutte, Imad El Haddad, Jakob Boyd Pernov, Hélène Paule Angot, Lubna Dada
Christophe Ballif, Luca Massimiliano Antognini, Samira Alexandra Frey, Mohammad Beygi
Julia Schmale, Ivo Fabio Beck, Benjamin Jérémy Laurent Heutte, Imad El Haddad, Jakob Boyd Pernov, Hélène Paule Angot, Lubna Dada