Êtes-vous un étudiant de l'EPFL à la recherche d'un projet de semestre?
Travaillez avec nous sur des projets en science des données et en visualisation, et déployez votre projet sous forme d'application sur Graph Search.
3D ultrasound (US) acquisition acquires volumetric images, thus alleviating a classical US imaging bottleneck that requires a highly-trained sonographer to operate the US probe. However, this opportunity has not been explored in practice, since 3D US machines are only suitable for hospital usage in terms of cost, size and power requirements. In this work we propose the first fully-digital, single-chip 3D US imager on FPGA. The proposed design is a complete processing pipeline that includes pre-processing, image reconstruction, and post-processing. It supports up to 1024 input channels, which matches or exceeds state of the art, in an unprecedented estimated power budget of 6.1 W. The imager exploits a highly scalable architecture which can be either downscaled for 2D imaging, or further upscaled on a larger FPGA. Our platform supports both real-time inputs over an optical cable, or test data feeds sent by a laptop running Matlab and custom tools over an Ethernet connection. Additionally, the design allows HDMI video output on a screen.
Mohamed Farhat, Davide Bernardo Preso, Ryan Holman
Romain Christophe Rémy Fleury, Janez Rus
Romain Christophe Rémy Fleury, Janez Rus