Publication

Qualification of a centerbody cavitation nuclei counter using optical techniques

1992
Conference paper
Abstract

A study about size measurements of cavitation nuclei is performed, by comparing results from different techniques as Holography, Phases Doppler and Centerbody Venturi. These systems are installed in line along a new installation, specially designed to guarantee optimal optical and flow conditions. First, to calibrate the different systems, pressure distribution inside the Centerbody Venturi is determined, using Laser velocimetry measurements. Response of the optical techniques is analyzed by injecting size calibrated latex particles. Then, the comparative measurements arc performed for different cavitation nuclei conditions.

About this result
This page is automatically generated and may contain information that is not correct, complete, up-to-date, or relevant to your search query. The same applies to every other page on this website. Please make sure to verify the information with EPFL's official sources.