Pulmonary hypertension is a hemodynamic disorder defined by an abnormal elevation of pulmonary artery pressure (PAP). Current options for measuring PAP are limited in clinical practice. The aim of this study was to evaluate if electrical impedance tomography (EIT), a radiation-free and non-invasive monitoring technique, can be used for the continuous, unsupervised and safe monitoring of PAP. In 30 healthy volunteers we induced gradual increases in systolic PAP (SPAP) by exposure to normobaric hypoxemia. At various stages of the protocol, the SPAP of the subjects was estimated by transthoracic echocardiography. In parallel, in the pulmonary vasculature, pulse wave velocity was estimated by EIT and calibrated to pressure units. Within-cohort agreement between both methods on SPAP estimation was assessed through Bland-Altman analysis and at subject level, with Pearson's correlation coefficient. There was good agreement between the two methods (inter-method difference not significant (P>0.05), bias +/- standard deviation of-0.1 +/- 4.5 mmHg) independently of the degree of PAP, from baseline oxygen saturation levels to profound hypoxemia. At subject level, the median per-subject agreement was 0.7 +/- 3.8 mmHg and Pearson's correlation coefficient 0.87 (P
, , , , ,
,
Nikolaos Stergiopulos, Georgios Rovas, Sokratis Anagnostopoulos, Vasiliki Bikia, Patrick Segers