Improvements in B(o) mapping and shimming were achieved by measuring the static field information in multiple subsequent echoes generated by an asymmetric echo-planar readout gradient train. With careful compensation, eddy current effects were shown to affect the adjustment of the shim coils minimally. In addition to reducing the time required for field mapping by two-fold, the sensitivity was simultaneously optimized irrespective of the prevalent T2/* present, thereby minimizing the error of the static field measurement to below 0.1 Hz. With adiabatic low flip-angle excitation, the time required for field mapping was below 1 second. (C) 2000 Wiley-Liss, Inc.
Jean-Philippe Thiran, Tobias Kober, Bénédicte Marie Maréchal, Jonas Richiardi
Fernando Porté Agel, Guiyue Duan, Daniele Gattari
Alain Nussbaumer, Heikki Tapani Remes, Abinab Niraula