Single-photon avalanche diode (SPAD) imagers can perform fast time-resolved imaging in a compact form factor, by exploiting the processing capability and speed of integrated CMOS electronics. Developments in SPAD imagers have recently made them compatible with widefield microscopy, thanks to array formats approaching one megapixel and sensitivity and noise levels approaching those of established technologies. In this paper, phasor-based FLIM is demonstrated with a gated binary 512×512 SPAD imager, which can operate with a gate length as short as 5.75 ns, a minimum gate step of 17.9 ps, and up to 98 kfps readout rate (1-bit frames). Lifetimes of ATTO 550 and Rhodamine 6G (R6G) solutions were measured across a 472×256 sub-array using phasor analysis, acquiring data by shifting a 13.1 ns gate window across the 50 ns laser period. The measurement accuracy obtained when employing such a scheme based on long, overlapping gates was validated by comparison with TCSPC measurements and fitting analysis results based on a standard Levenberg-Marquardt algorithm (
Giovanni Boero, Nergiz Sahin Solmaz, Reza Farsi
Mihai Adrian Ionescu, Junrui Zhang, Francesco Bellando, Pierpaolo Palestri, Luca Selmi