Are you an EPFL student looking for a semester project?
Work with us on data science and visualisation projects, and deploy your project as an app on top of Graph Search.
to follow up patients who inject themselves insulin with injection pen is a quite complex task. Due to the repetition of injections, and the numerous influent parameter the patient can easily made an injection mistake and it is difficult for health-takers to follow these injections and can detect errors in order to compensate this problem, a device which measures the volume of drug by using electric fields has been developed. A theoretical model, simulations and experiments have been conducted to validate this capacitive measurement technique. The experiments confirm Theory and simulations results. A sensitivity around 5117 for 10uL have been found and a good repeatability between three caps have been achieved which is equal to an error of 0.44fF for lOuL of insulin and correspond to an error of +/- 4.12%.
, ,
Katrin Beyer, Yves Weinand, Julien Gamerro, Savvas Saloustros, Andrea Settimi, Andrea Cabriada Ascencio