Publication

Real-Time High-Sensitivity Impedance Measurement Interface for Tethered BLM Biosensor Arrays

Abstract

This paper presents a switched-capacitor (SC) current integrator circuit for impedance measurement of tethered bilayer lipid membrane (tBLM) biosensors. The circuit comprises a small number of high performance components enabling enhanced experimental flexibility and reliability. The sensitivity is improved significantly by suppressing the output offset through pseudo-differential operation, using R-C components for the reference impedance. The sensing and reference electrodes are excited with low-amplitude differential voltage pulses and the current response to membrane resistance (RM) change of the tBLM biosensor is converted to voltage by a precision, low-noise SC integrator available as a single-package IC. Tests with both electrical models and actual biosensors demonstrated that the proposed circuit operates with high sensitivity and can be used in single chip versions for low-cost and high-sensitive tBLM biosensor arrays, featuring multiple electrode sites.

About this result
This page is automatically generated and may contain information that is not correct, complete, up-to-date, or relevant to your search query. The same applies to every other page on this website. Please make sure to verify the information with EPFL's official sources.