Publication

Modeling of a capacitive sensor dedicated to drug injection

Catherine Dehollain
2018
Conference paper
Abstract

to follow up patients who inject themselves insulin with injection pen is a quite complex task. Due to the repetition of injections, and the numerous influent parameter the patient can easily made an injection mistake and it is difficult for health-takers to follow these injections and can detect errors in order to compensate this problem, a device which measures the volume of drug by using electric fields has been developed. A theoretical model, simulations and experiments have been conducted to validate this capacitive measurement technique. The experiments confirm Theory and simulations results. A sensitivity around 5117 for 10uL have been found and a good repeatability between three caps have been achieved which is equal to an error of 0.44fF for lOuL of insulin and correspond to an error of +/- 4.12%.

About this result
This page is automatically generated and may contain information that is not correct, complete, up-to-date, or relevant to your search query. The same applies to every other page on this website. Please make sure to verify the information with EPFL's official sources.