Publication

CMOS image sensors for subretinal implant system

Mihai Adrian Ionescu, Philippe Renaud, Marco Mazza
2003
Article de conférence
Résumé

Visual capabilities recovery for some kind of illness is possible through subretinal implantable device stimulation. Two possible approaches for retinal pixel are proposed, fabricated in 0.35mum CMOS, tested and compared including electronic response on a human tissue like interface. In these solutions, power consumption has been proved to be dominated by electrode stimulation and, in normal condition, typical consumption is below 1 mW. The proposed pixel architectures, with 120 mum (LICOS) and 100 mum (WATCH) inter-pixel step, provide the smallest, standalone, highly integrated, subretinal implantable chip ever reported.

À propos de ce résultat
Cette page est générée automatiquement et peut contenir des informations qui ne sont pas correctes, complètes, à jour ou pertinentes par rapport à votre recherche. Il en va de même pour toutes les autres pages de ce site. Veillez à vérifier les informations auprès des sources officielles de l'EPFL.

Graph Chatbot

Chattez avec Graph Search

Posez n’importe quelle question sur les cours, conférences, exercices, recherches, actualités, etc. de l’EPFL ou essayez les exemples de questions ci-dessous.

AVERTISSEMENT : Le chatbot Graph n'est pas programmé pour fournir des réponses explicites ou catégoriques à vos questions. Il transforme plutôt vos questions en demandes API qui sont distribuées aux différents services informatiques officiellement administrés par l'EPFL. Son but est uniquement de collecter et de recommander des références pertinentes à des contenus que vous pouvez explorer pour vous aider à répondre à vos questions.